Cobalt- Chrome
Scope
This information is representative of cobalt-chromium-molybdenum
alloy high-strength forgings for hip prosthesis. The material is comparable to Specification ASTM-F75
and F-799.
The reported properties are an average of all independent
laboratory testing and represents a minimum of 500 tests.
Significance and Use
The purpose of this information is to identify a low cost
alternative to forged CoCrMo product.
Mechanical Requirements
The data in Table's I through 4 compares the results of the Comtech
Casting to the requirements of ASTM-F75 and F799. Grain size's
are a minimum of five or finer.
Rationale
The Comtech Casting is the only alternative to a traditional
forged product from wrought material currently. The reported
tensile strengths are comparable with the minimum requirements
of the wrought and forged specifications ASTM-F1537 and F799.
The resulting microstructure is far superior in both carbide distribution and grain size.
These results were achieved without incorporating hot isostatic
pressing or solution annealing.
Rotating Fatigue Testing
|
|
|
|
|